Microprocessor controlled KARL FISCHER
Type-Karl Fischer Moisture Titrator
Model- MKV-710D
Standard configuration MCU-710D+MKV-710+Automatic Solvent Change unit
Measuring method- Karl Fischer Volumetric titration
Measuring range-1)Water content : 0.1 to 500mgH2O (depends on KF reagent factor)
2) Concentration : 1ppm to 100% H2O
Burette precision-10ml burette +/- 0.015ml ; Reproducibility +/- 0.005mL
Endpoint detection By polarized potential level detected with a twin platinum electrode
Other Description-EP sense method Detection of potential level maintained during preset end time
End time range :-1 to 99 Second
Titration form Normal titration / Back titration (Option additional burette required)
Required solvent Volume-30 to 100ml(in S-type titration vessel)
Key operation-Touch panel
Displays Lcd Dimension- 8.4-inch color LCD 800 x 600 dots

Language of the Display- English / Japanese / Mandarin Chinese
Display Type-Simultaneous 2-channel display
Calculation Concentration of water content-statistics data processing (mean , SD and RSD) and automatic averaging of blank value and factor value
Data storage Method:-120 (per channel)
Measurement results:-5,000 samples (per channel)
Operation history: -100,000 (Total for 2 channels)
DI support function Access Restrictions:-Operator authentication by operator ID and password
Rights Administration:- Each operator is assigned to a specific right group
Audit Trail: Operation history is automatically recorded
Electronic Approval / Electronic Signature:- Electronic approval and signature based on the workflow
Backup / Restoration: -Automatic backup with the predetermined period
Archive / Archive View:-Data storage. Archived data can be viewed on the main control unit
Data Capacity:-The remaining capacity can be checked any time via an icon on the main screen
Security Setup:-Password length and lockout settings can be set
LIMS Connection:-Data output to LIMS systems
External I/O- RS-232C Port x4 :-Dot matrix printer , Electronic balance , Data Capture Software (SOFT-CAP) , Evaporator
USB ×1 :- USB flash drive , A4 printer , Thermal printer , Keyboard , Barcode reader , Foot switch , USB HUB
LAN ×1 :-Personal computer (PC)
Extensibility Measuring instrument :-Karl Fischer Moisture Titrator (MKV-710/MKC-710 (One of those instruments can be added)
Automatic piston burette :Can control max 2 burette drives (Including two built-in burette drives)
Evaporator -ADP-611
Ambient condition-1) Temperature : 5 to 35°C
2) Humidity :85%RH or below (no condensation)
Power source-AC100 to 240V +/- 10% 50/60 Hz
Power consumption Main unit :-Approx. 30W
Dimensions Touch panel controller : -225(W) x 190(D) x 42(H) mm
Titration unit : 141(W) x 292(D) x 367(H) mm (not incl. tubing)
Stirrer :-107(W) x 206(D) x 322(H) mm (not incl. Solvent Change unit)
Solvent Change Unit :- 240(W) x 140(D) x 400(H) mm (not incl. tubing)
Weight Touch panel controller : -Approx.1.5kg
Titration unit :-Approx.4.0kg
Stirrer : -Approx.2.0kg
Solvent Change Unit :-Aprox.0.6kg
Conformity standard CE marking
EMC: - EN61326-1